Automatic Glycohemoglobin Analyzer ADAMS A1c HA-8180T
The ADAMS A1c, HA-8180T is a fully automated Glycohemoglobin analyzer based on HPLC (High Performance Liquid Chromatography) and is highly effective for ß-thalassemia testing.- Measurement speed: 3.5 minutes/test
- Measurement Item as below
- HbA2 HbF
- HbA1c (Stable HbA1c, S-A1c),
- HbS, HbC, HbE and HbD can be detected.
- Column design with no more pre-filter change.
- Color LCD with reagent level, graphs on screen.
Specifications
Samples | Whole blood or hemolysis sample |
Measurement items | HbA1c(Stable HbA1c, S-A1c), HbA2 and HbF (HbS, HbC, HbE and HbD can be detected.) |
Reagents *1 | ELUENT 80A, ELUENT 80B, ELUENT 80CT and HEMOLYSIS WASHING SOLUTION 80H |
Column *2 | COLUMN UNIT 80T |
Measurement principle | Reversed-phase cation exchange chromatography |
Detection method | Dual-wavelength colorimetry (wavelengths measured: 420nm/500nm) |
Resolution | 0.1% Ratio, 1 mmol/mol |
Measurement range *3 | HbA1c: 3-20%, 9-195 mmol/mol HbA2: 2-10% HbF: 0-100% |
Guaranteed measurement range *4 | HbA1c: 4.2-12.9%, 22-117 mmol/mol HbA2: 2.4-6.3% HbF: 0.1-14.3% |
Processing speed | 210 seconds per sample |
Sample consumption | Approx.14μL(whole blood) |
Sample container | Cap-Piercing |
Rack type | ARKRAY racks |
Number of measurement samples | One-way transportation (factory set): Maximum 50 samples Loop transportation: Maximum 100 samples |
Column temperature | Approx. 39℃ |
Display | Color graphic LCD with backlight |
Memory capacity | 500 test results ( including calibration results ) |
Built-in Printer | Termal printer, 58mm thermal paper |
External output | Serial 1 port (Can be optionally used as an Ethernet port.) |
Measurement conditions | Temperature: 15-30℃ Humidity 20-80% relative humidity (noncondensing ) |
Required sample volume | Blood collection tube: Minimum 10mm away from the bottom Sample cup: 400μL or more |
Warm-up time | Maximum 30 min. |
Dimensions | 530(W)×530(D)×530(H)mm (Not including the hemolysis washing solution bottle) |
Weight | Analyzer: Approx. 39kg, Sampler: Approx. 4kg |
Power supply | AC100V-240V ±10% 50/60Hz |
Power consumption | Max. 300 VA |
*1, *2 Reagents and column are sold separately from HA-8180T analyzer.
*3 Error-free measurement ranges.
*4 Guaranteed ranges for obtaining results equivalent to those from HA8160 TP mode measurement.